Channel selector for optimization of test and callibration procedures of ICTM pressure sensors
Conference object (Published version)
Metadata
Show full item recordAbstract
This paper presents channel selector developed for improvement of the efficiency of the test and callibration
procedure of ICTM pressure sensors. So far, this procedure allowed testing of only one sensor at the time. The channel
selector enables simultaneous testing of up to five pressure sensors. This implies many benefits, such as: reduction of
power consumption, reduction of the amount of gas used in the measurement system, significant saving of time needed
for testing of a small series of pressure transmitters. Realization of the channel selector is based on use of a reed relay
which is driven by specially designed electronics. There are two modes of operation of this device. The first one is by
manual channel selection, while the second one is based on PC control realized by using dedicated software and
microcontroller which are incorporated in the device itself and connected via USB communication. During the test and
callibration procedure pressure is generated by MENSOR ...APC-600 pressure calibrator, while temperature is
controlled using temperature test chamber HERAEUS VÖTSCH VMT 08/140.
Keywords:
pressure sensor / callibration / channel selector / reed relay / USB communicationSource:
Proceedings - 7th International Scientific Conference on Defensive Technologies, OTEH 2016, October 6-7, Belgrade, Serbia, 2016, 377-380Publisher:
- Belgrade : Military Technical Institute
Funding / projects:
- Micro- Nanosystems and Sensors for Electric Power and Process Industry and Environmental Protection (RS-32008)
- Measurements in a smart grid concept (RS-32019)
Collections
Institution/Community
IHTMTY - CONF AU - Poljak, Predrag AU - Vorkapić, Miloš AU - Randjelović, Danijela PY - 2016 UR - https://cer.ihtm.bg.ac.rs/handle/123456789/3896 AB - This paper presents channel selector developed for improvement of the efficiency of the test and callibration procedure of ICTM pressure sensors. So far, this procedure allowed testing of only one sensor at the time. The channel selector enables simultaneous testing of up to five pressure sensors. This implies many benefits, such as: reduction of power consumption, reduction of the amount of gas used in the measurement system, significant saving of time needed for testing of a small series of pressure transmitters. Realization of the channel selector is based on use of a reed relay which is driven by specially designed electronics. There are two modes of operation of this device. The first one is by manual channel selection, while the second one is based on PC control realized by using dedicated software and microcontroller which are incorporated in the device itself and connected via USB communication. During the test and callibration procedure pressure is generated by MENSOR APC-600 pressure calibrator, while temperature is controlled using temperature test chamber HERAEUS VÖTSCH VMT 08/140. PB - Belgrade : Military Technical Institute C3 - Proceedings - 7th International Scientific Conference on Defensive Technologies, OTEH 2016, October 6-7, Belgrade, Serbia T1 - Channel selector for optimization of test and callibration procedures of ICTM pressure sensors SP - 377 EP - 380 UR - https://hdl.handle.net/21.15107/rcub_cer_3896 ER -
@conference{ author = "Poljak, Predrag and Vorkapić, Miloš and Randjelović, Danijela", year = "2016", abstract = "This paper presents channel selector developed for improvement of the efficiency of the test and callibration procedure of ICTM pressure sensors. So far, this procedure allowed testing of only one sensor at the time. The channel selector enables simultaneous testing of up to five pressure sensors. This implies many benefits, such as: reduction of power consumption, reduction of the amount of gas used in the measurement system, significant saving of time needed for testing of a small series of pressure transmitters. Realization of the channel selector is based on use of a reed relay which is driven by specially designed electronics. There are two modes of operation of this device. The first one is by manual channel selection, while the second one is based on PC control realized by using dedicated software and microcontroller which are incorporated in the device itself and connected via USB communication. During the test and callibration procedure pressure is generated by MENSOR APC-600 pressure calibrator, while temperature is controlled using temperature test chamber HERAEUS VÖTSCH VMT 08/140.", publisher = "Belgrade : Military Technical Institute", journal = "Proceedings - 7th International Scientific Conference on Defensive Technologies, OTEH 2016, October 6-7, Belgrade, Serbia", title = "Channel selector for optimization of test and callibration procedures of ICTM pressure sensors", pages = "377-380", url = "https://hdl.handle.net/21.15107/rcub_cer_3896" }
Poljak, P., Vorkapić, M.,& Randjelović, D.. (2016). Channel selector for optimization of test and callibration procedures of ICTM pressure sensors. in Proceedings - 7th International Scientific Conference on Defensive Technologies, OTEH 2016, October 6-7, Belgrade, Serbia Belgrade : Military Technical Institute., 377-380. https://hdl.handle.net/21.15107/rcub_cer_3896
Poljak P, Vorkapić M, Randjelović D. Channel selector for optimization of test and callibration procedures of ICTM pressure sensors. in Proceedings - 7th International Scientific Conference on Defensive Technologies, OTEH 2016, October 6-7, Belgrade, Serbia. 2016;:377-380. https://hdl.handle.net/21.15107/rcub_cer_3896 .
Poljak, Predrag, Vorkapić, Miloš, Randjelović, Danijela, "Channel selector for optimization of test and callibration procedures of ICTM pressure sensors" in Proceedings - 7th International Scientific Conference on Defensive Technologies, OTEH 2016, October 6-7, Belgrade, Serbia (2016):377-380, https://hdl.handle.net/21.15107/rcub_cer_3896 .