@article{
author = "Sejmanović, Dragana M. and Petković, Branka B. and Budimir, Milana V. and Sovilj, Sofija P. and Jovanović, Vladislava M.",
year = "2011",
abstract = "An all solid-state Ag(I) ion-selective electrode has been prepared by simply immersing a glassy carbon rod coated with PVCAc, which contained plasticizer and additive but no ionophore, into the AgNO3 solution. The response of the electrode was linear with a Nernstian slope of 60.25 mV/decade within the concentration range from 1 x 10(-1) to 1 x 10(-5) M and with a detection limit of 4.25 x 10(-6)M. The stability as an effect of various cations was defined. The electrode is suitable for use in high acidic solutions (pH LT 1 to 7) and has successfully been applied for the determination of silver(I) concentrations in different samples.",
publisher = "Wiley-Blackwell, Malden",
journal = "Electroanalysis",
title = "Characterization of a Silver Modified PVCAc Electrode and Its Application as a Ag(I)-Selective Potentiometric Sensor",
volume = "23",
number = "8",
pages = "1849-1855",
doi = "10.1002/elan.201000745"
}