@article{
author = "Drljevic-Djuric, K. M. and Jović, Vladimir D. and Lačnjevac, Uroš and Avramov Ivić, Milka and Petrović, Slobodan D. and Mijin, Dušan and Djordjevic, S. B.",
year = "2010",
abstract = "The oxidative behavior of antibiotic roxithromycin standard was studied at a gold electrode in 0.05 M NaHCO3 using cyclic linear sweep voltammetry and differential pulse voltammetry. It was found that the value of the oxidative peak of pure roxithromycin at 0.81 V vs. SCE in 0.05 M NaHCO3 at a scan rate of 50 mV s(-1) is a linear function of the concentration in a range 0.10006-0.654 mg cm(-3). It was also found that peak current density at 0.75 V vs. SCE at a scan rate of 2 mVs(-1), pulse amplitude of 25 mV and pulse time of 0.1 s exhibits linear dependence on the concentration of roxithromycin from 0.1006 to 0.476 mg cm(-3). Roxithromycin as a content of solid dosage form and urine was quantitatively determined and the obtained results were checked by HPLC.",
publisher = "Oxford : Pergamon-Elsevier Science Ltd",
journal = "Electrochimica Acta",
title = "Voltammetric and differential pulse determination of roxithromycin",
volume = "56",
number = "1",
pages = "47-52",
doi = "10.1016/j.electacta.2010.09.067"
}